Alle Kategorien
Ihr Warenkorb
VICTOR 26/26H Multifunction Process Calibrator
Artikel-ID #26
Listenpreis: $1850.00
Preis: $1400.00
Mengenbereich (Einheit) Preis (pro Einheit) Rabatt
5 $1250.00 $150.00
10 $1150.00 $250.00
50 $1100.00 $300.00
Menge: Einheit(s)
Beschreibung:

VICTOR26 Multifunction Process Calibrator Signal Generator With Accuracy Of 0.01% 5 Digits Display 24V Loop Circuit IP 65 Protection Grade


* Basic feature
* A multifunctional on-spot process calibrator with high accuracy and high stability, which can provide powerful measurable
functions, output and calibrates all parameters for all kinds of process.
* Basic feature of input measurement
* The max accuracy of measurement can reach 0.01%, and displays as 5 digits.
* Measurement functions include: DC voltage, DC current, resistance, RTD, TC, frequency, pulse counting, switch volume, pressure,and pressure leakage ect.
* Undertake two-wired, three-wired and four-wired resistance and RTD measurement.
* Switch test of fast response allows capturing the process parameters for switching actions.
* Data settle function: data-holding, zero, proportion, mean, Max/min value ect.
* Built-in 24V loop circuit and loop current measurement
* Basic feature of analog output
* Max accuracy of output can reach 0.01%, and output displays as 6 digits.
* Output functions include: DC voltage, DC current, analog resistance, RTD, TC, analog transmitter output, frequency,
pulse counting, switch volume, and pressure output ect.
* Output augment and decrease setting, and can connect or disconnect output, which is easy for operation.

DCI output manual-step, auto-step and auto-ramping output functions.
Built-in 24V loop circuit.
* Two independent channels for measurement and input, allowing in-time measurement and outputs of process signals simultaneously;
multiple data display to display input measurement value and output settings.
* Facilitate pressure calibration by external a series of VPM-S digital pressure module through serial RS-232 injection (LEMO
plug). There are abundant varieties and range specification of pressure modules to meet any pressure measurement needs. And the
basic accuracy can reach 0.05%, and a variety of pressure units are available for pressure calibration.
* Provide varieties of selectable temperature graduation. Built-in accurate temperature transmitter, with high accurate TC
cold-compensation function, and RTD cold compensation can be selected both for connection or disconnection.
* PC on-line communication function

* The meter provides two ways to communicate with the user’s upper PC and can communicate by the infrared isolated serialcommutation module of the compatible USB interface.
* Supports SCPI commands of industrial standard, and can realize remote control of the meter by equipment driving software.
* Note: the PC communication function is not available when the meter is communicating with the pressure module in HART communication function.
* The meter is equipped with data storage with large capacity, to store as many as 1000(groups) data of field calibration. The data can be read from the meter or uploaded to the user’s PC.
* The meter employs panel calibration technique, connects to standard device, undertakes panel calibration operation according to set range, and saves relevant calibrated data, fulfills the periodic calibration work stipulated by the meter to ensure the accuracy and performance.
* The meter employs the navigation key to make increment or decrement of the setting value, which makes operation convenient and easy.
* 3.2 inches TFT display screen, supporting multiple data display and intuitive interface.
* Other functions: auto power-off, VF, auto turn- off for backlight, setting of backlight, data and time display and setting,
ambient temperature detection, power frequency setting ect.
* two ways of power supply
* The instrument can supply power through the four 5th alkaline batteries in the battery case, or use the adaptor to supply.

Specification
measurement function
range
measurement range
resolution
accuracy
remark
DC voltage
DCV
50mV
-5.000mV~
55.000mV
1μV
0.01+0.01
Input resistance: about 100MΩ
500mV
-50.00mV~
550.00mV
10μV
0.01+0.01
5V
-0.5000V~
5.5000V
0.1mV
0.01+0.01
Input resistance: about 1MΩ
30V
-5.000V~
35.000V
1mV
0.01+0.01
DC current
DCI
50mA
-5.000mA~
55.000mA
1μA
0.01+0.01
shunt resistance: about 10Ω
Ohms
OHM
(4W)
500Ω
0.00Ω~550.00Ω
0.01Ω
0.01+0.01
500Ω about 1mA simulation
5KΩ about 0.1mA simulation
open-loop voltage: about 2.5V
wire resistance is excluded in the accuracy
5KΩ
0.0000 KΩ~
5.5000KΩ
0.1Ω
0.01+0.01
frequency
FREQ
50KHz
3Hz~50.00000KHz
0.01Hz
0.01+0.00004
input resistance: more than 100 kΩ
frequency sensitivity:
square wave is 3V Vp-p at least
50% duty ratio
CPM
180~3000000 CPM
1CPM
±2个字
Thermal couple
TC
R*
0°C~1767°C
1°C
0~500℃ : 1.8℃
500~1767℃ : 1.5℃
Employ ITS-90 temperature standard
Errors for cold-end compensation is excluded in the accuracy
S*
0°C~1767°C
K
-100.0°~1372.0°C
0.1°C
-100.0~0.0℃ : 1.2℃
0.0~1372.0℃ : 0.8℃
E
-50.0°C~1000.0°C
-50.0℃~0.0℃ : 0.9℃
0.0~1000.0℃ : 1.5℃
J
-60.0°C~1200.0°C
-60.0~0.0℃ : 1.0℃
0.0~1200.0℃ : 0.7℃
T
-100.0°C~400.0°C
-100.0~0.0℃ : 1.0℃
0.0~400.0℃ : 0.7℃
N
-200.0°~1300.0°C
-200.0~0.0℃ : 1.5℃
0.0~1300.0℃ : 0.9℃
B*
600°C~1820°C
1°C
600~800℃ : 2.2℃
800~1000℃ : 1.8℃
1000~1820℃ : 1.4℃
L
-60.0°C~900.0°C
0.1°C
-60.0~0.0℃ 0.7℃
0.0~900.0℃ 0.5℃
U
-100.0°C~600.0°C
0.1°C
-100.0~0.0℃ 0.7℃
0.0~600.0℃ 0.5℃
Resistance temperature
RTD
Pt100
385
-200.0°C~800.0°C
0.1°C
-200.0~0.0℃ : 0.5℃
0.0~400.0℃ : 0.7℃
400.0~800.0℃ : 0.8℃
Employ ITS-90 temperature standard
Errors due to the mismatch of lead resistance for lead resistance, 3-wired or 4-wired measurements are excluded in the accuracy
Pt1000
385
-200.0°C~630.0°C
-200.0~100.0℃: 0.3℃
100.0~300.0℃ : 0.5℃
300.0~630.0℃ : 0.7℃
Pt200
385
-200.0°C~630.0°C
-200.0~100.0℃: 0.8℃
100.0~300.0℃ : 0.9℃
300.0~630.0℃ : 1.0℃
Pt500
385
-200.0°C~630.0°C
-200.0~100.0℃: 0.4℃
100.0~300.0℃ : 0.5℃
300.0~630.0℃ : 0.7℃
Cu10
-100.0°C~260.0°C
1.8℃
Cu50
-50.0°C~150.0°C
0.7℃
Pulse counting
PULSE
100000 cycles
1~100000cycles
1cycle
±2个字
see the related indexes for frequency
switch volume measurement
SWITCH
 
CLOSE / OPEN
 
 
about 1mA simulation
short circuit displays CLOSE
open circuit displays OPEN
the threshold value is about 200~300Ω
Continuity test
CONNECT
500Ω
≤ 50Ω vocalize
0.01Ω
 
500Ω
about 1mA simulation
Loop supply
LOOP
24 V
 
 
10%
short circuit protection
Max circuit : 22 mA
Max input voltage: 30 V DC
output resistance: 250Ω nominal value

Monkey Lin
Lin
Lin
Lin
Xia Lin