Description:

VICTOR 01S Temperature Source highest output precision is 0.05% and the output display is 5 digits Multifunction Calibrator

  Basic characteristics of the product

1. Analog output characteristics
l Basic functions
DC voltage DCV, resistance OHM, thermocouple TC, thermal resistance RTD
l The maximum output accuracy is 0.05% and the output display is 5 digits.
l Output settings can be manually increased or decreased by position.
l H/L settings are available for OHM excitation current.
l The output temperature value and the corresponding temperature signal are displayed on the same screen.
2. It has a built-in high-precision temperature sensor for high-precision cold-end compensation.
Thermocouple cold-end compensation can be set, including manual compensation, automatic compensation, compensation access disconnection, etc.
3. It has maintenance setup functions, including
Auto power-off, backlight, flashlight, etc.
4. Other
l The segment code is displayed on an LCD with a white backlight.
l It has a LED flashlight for lighting.
l It is powered by 3 AA battery cells. The battery compartment door is convenient for battery replacement. Low-battery detection is available.
l It is small, firm, reliable, suitable for field use, and cheap.
l User-optional large-suction magnetic sling is provided.
Basic function and performance of the product (technical index)
 
Analog output function [Applied within one year after calibration, 23℃±5℃, 20-70% RH, accuracy=± (% set value +% reading)]  
Output function Range Output range Resolution Accuracy Remarks:
DC voltage
DCV
100mV -10.000mV~110.00mV 0.01mV 0.05%+0.03mV Maximum output current: 1mA
1V -0.1000V~1.1000V 0.0001V 0.05%+0.3mV Maximum output current: 5mA
Ohm
OHM
400Ω 0~400.0Ω 0.1Ω 0.05%+0.2Ω The excitation current is ±0.1~±5mA
Maximum output voltage 2V
When the excitation current is ±0.1~0.5mA, additional error of 0.1Ω is added
The accuracy does not include lead resistance
4KΩ 0~4.000KΩ 0.05%+2Ω
Thermocouple
TC
R 0°C~1767°C 1°C 0~100℃
10~1767℃
0~100℃
100~1767℃
1.5℃
1.2℃
1.5℃
1.2℃
Use ITS-90 temperature scale
The accuracy does not include the error of cold junction compensation
The accuracy does not include sensor inaccuracy
The accuracy does not include the influence of thermoelectric potential
R: Platinum Rhodium 13- Platinum
S: Platinum Rhodium 10- Platinum
K: Nickel chromium - nickel silicon
E: Nickel chromium - copper nickel (constantan)
J: Iron - copper nickel (constantan)
T: Copper - nickel (constantan)
N: Nickel chromium silicon nickel silicon
B: Platinum Rhodium 30 - Platinum
A: Tungsten rhenium 5- Tungsten rhenium 20
C: Tungsten rhenium 5- Tungsten rhenium 26
D: Tungsten rhenium 3- Tungsten rhenium 25
S 0°C~1767°C
K -200.0°C~1372.0°C 0.1°C -200.0~-100.0℃
-100.0~400.0℃
400.0~1200.0℃
1200.0~1372.0℃
0.6℃
0.5℃
0.7℃
0.9℃
E -200.0°C~1000.0°C -200.0~-100.0℃
-100.0~600.0℃
600.0~1000.0℃
0.6℃
0.5℃
0.4℃
J -200.0°C~1200.0°C -200.0~-100.0℃
-100.0~800.0℃
800.0~1200.0℃
0.6℃
0.5℃
0.7℃
T -250.0°C~400.0°C -250.0~400.0℃ 0.6℃
N -200.0°C~1300.0°C -200.0~-100.0℃
-100.0~900.0℃
900.0~1300.0℃
1.0℃
0.7℃
0.8℃
B 600°C~1820°C 1°C 600~800℃
800~1820℃
1.5℃
1.1℃
A 0°C~2500°C 1°C 0~1600℃
1600~2000℃
2000~2500℃
    2.0℃
    2.2℃
    2.4℃
C 0°C~2310°C 1°C 0~1600℃
1600~2000℃
2000~2310℃
    2.0℃
    2.2℃
    2.4℃
D 0°C~2310°C 1°C 0~100℃
100~270℃
270~1200℃
270~2310℃
    2.6℃
    2.4℃
    2.2℃
    2.4℃
Thermal resistance
RTD
Pt100
385
-200.0°C~800.0°C 0.1°C -200.0~0.0℃
0.0~400.0℃
400.0~800.0℃
0.3℃
0.5℃
0.8℃
Use Pt (385) temperature scale
The excitation current is ±0.1~±5mA
Maximum output voltage 2V
When the excitation current is ±0.1~0.5mA, additional error of 0.1Ω is added
The accuracy does not include lead resistance
The accuracy does not include the influence of thermoelectric potential
Support pulse transmitters and PLCs with pulse times as short as 10ms
Pt200
385
-200.0°C~630.0°C   -200.0~100.0℃
100.0~300.0℃
300.0~630.0℃
0.8℃
0.9℃
1.0℃
Pt500
385
-200.0°C~630.0°C -200.0~100.0℃
100.0~300.0℃
300.0~630.0℃
0.4℃
0.5℃
0.7℃
Pt1000
385
-200.0°C~630.0°C -200.0~100.0℃
100.0~300.0℃
300.0~630.0℃
0.2℃
0.5℃
0.7℃
Cu50 -50.0°C~150.0°C 0.6℃
 
 
Other characteristics:
l Uncertainty includes standard uncertainty, hysteresis, nonlinearity, repeatability, and typical long-term stability over the period mentioned (K = 2).
l Maximum applied voltage at input end: about 30Vpk; Maximum applied current at input end: about 25mA
l Load characteristics: capacitive load ≥0.01uF (DCV/OHM/TC/RTD/FREQ); Inductive load ≥ 0.01uF 100uH (DCA)
Output load affects DCmV (0.001% full scale + 1nV)/mA; DCV (0.001% full scale + 1nV)/mA
l The temperature range of internal temperature compensated sensor RJC is -10 to 50°C. The temperature measurement accuracy at 18 to 28°C is ±0.5°C, and the temperature measurement accuracy at other temperature is ±1°C. Cold end compensation time is 10S/ times.
l Temperature coefficient: 0.1 × basic accuracy/℃ (temperature range < 18℃ or > 28℃)


III.General characteristics of the product
l Operating temperature and humidity: 0 to 50ºC  ≤80%RH without condensation; 40 to 50ºC  ≤70%RH.
l Storage temperature and humidity: -25 to 60ºC  ≤90%RH without condensation.
l Electrical safety: EN61010-1:2001
Withstand voltage: AC3540V(50/60Hz)/5 seconds between terminal and housing.
Insulation impedance: DC1000V/100MΩ or more between terminal and housing.
l Electromagnetic compatibility (EMC): EN61326-1:2006
Performance criterion 2 is met, i.e. the function and performance are temporarily reduced or lost, but can be recovered by themselves.
l Protection grade IP40.
l Vibration and fall: IEC 60068-2-64:2008, IEC 60068-2-32:2008
Random, 2g, 5-500Hz; 1m drop test.
l CE certification.
l Quality standard: developed, designed and manufactured in accordance with TUV ISO 9001.
l Calibration period and preheating time:
To ensure the accuracy of the Table, the calibration period is one year.
The startup preheating time is more than 30 minutes.
l Instrument display and key
Segment code LCD display, white backlight
Rubber key is adopted
l Instrument power supply
3 pcs of AA alkaline batteries for power supply
l Dimension and weight
 
IV.Peripheral configuration of the product
l Standard configuration
         Industrial test conductor: 1 pair
User’s Manual: 1 manual
AA alkaline battery: 3 pcs
Graphic carton: 1 item
Thermocouple adapter: 1 item
 

Monkey Lin
Lin
Lin
Lin
Xia Lin