Qty.Range(unit) | Price(per unit) | Discount |
100 | $1199.99 | $600.00 |
VICTOR 25 Multi-function Process Calibrator
Detailed Product Description
Measuring function |
Range |
Measuring range |
Resolving power |
Accuracy |
Remark |
DC Voltage |
50mV |
-5.000mV55.000mV |
1μV |
0.02%+10 |
input impedance: about100MΩ |
500mV |
-10.00mV550.00mV |
10μV |
0.02%+2 |
||
5V |
-0.1000V5.5000V |
0.1mV |
0.02%+5 |
input impedance: 1MΩ |
|
50V |
0.00V55.00V |
10mV |
0.03%+5 |
||
DC Current |
50mA |
-5.000mA55.000mA |
1μA |
0.02%+2 |
input impedance: 10Ω |
OHM |
500Ω About 1mA exciting |
0.00Ω550.00Ω |
0.01Ω |
0.02%+0.1Ω |
Plough voltage:about 2.5V not including the accuracy of down-lead resistance |
5KΩ About 0.1mA exciting |
0.0000 KΩ5.5000KΩ |
0.1Ω |
0.02%+0.5Ω |
||
Frequency |
100Hz |
2Hz110.0Hz |
0.1Hz |
±2 |
input impedance: 100 kΩ or more Square wave at least 1Vp-p 50% duty cycle |
1KHz |
0.100KHz1.100KHz |
1Hz |
|||
10KHz |
0.10KHz11.0KHz |
0. 1KHz |
|||
Thermocouple |
R |
-40°C1760°C |
1°C |
-4500 : 1.8 501760 : 1.5 |
Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error |
S |
-20°C1760°C |
||||
K |
-200°C950°C |
0.1°C |
-2000 : 1.2 01370 : 0.8 |
||
E |
-200°C500°C |
-2000 : 0.9 01000 : 1.5 |
|||
J |
-200°C700°C |
-2000 : 1.0 01200 : 0.7 |
|||
T |
-200°C400°C |
-2000 : 1.0 0400 : 0.7 |
|||
N |
-200°C1000°C |
-2000 : 1.5 01300 : 0.9 |
|||
B |
400°C1800°C |
1°C |
400800 : 2.2 8001000 : 1.8 10001800 : 1.4 |
||
Thermo resistance |
PT100 |
-200°C700°C |
0.1°C |
-2000 : 0.5 0400 : 0.7 400850 : 0.8 |
PT100-385 thermometric scale not including the accuracy of down-lead resistance |
PT1000 |
-200°C630°C |
-200100 : 0.22 100300 : 0.3 300630 : 0.4 |
|||
Cu50 |
-50°C150°C |
-500 : 0.5 0150 : 0.7 |
|||
Switch mete measure |
About 1mA exciting |
CLOSE / OPEN |
|
|
|
Continuity test |
500Ω About 1mA exciting |
≤50Ω vocalism |
0.01Ω |
|
|