Qty.Range(unit) | Price(per unit) | Discount |
50 | $180.00 | $40.00 |
Maximum voltage between any terminal and earth ground. | 1000V dc or ac RMS voltage |
Legal Compliance | Complies with IEC61010.1-2001 to 300V Over voltage IV Pollution Degree 2 (Safety Standard issued by IEC) |
Fuse Protection for mA or µA inputs Fuse Protection for A input |
0.5A 250V FAST Fuse 10A 250V FAST Fuse |
Identification tags | CE |
Display(VFD) |
Digital: 55000 counts primary display; 5500 counts secondary display; updates 5/second(slow), 20/second(fast). |
Operating Temperature | 0~50℃ |
Storage Temperature | -10~55℃ |
Relative Humidity |
0℃~30℃ ≤75% 30℃~40℃ ≤50% |
Altitude | 0-2000m(according to IEC61010 CAT.Ⅱ,300V) |
Power | 250V±10%AC,power frequency:50Hz or 60Hz,power: |
Consumption | About 5W |
Temperature coefficient | 0.15 × (specified accuracy)% / °C, range<18 °C or >28 °C ) |
EMC | Complies with IEC61326-1,Group 1,Class B |
Size | 245 × 220 × 82(mm) |
Weight | About 2000g |
Calibration Interval | One Year |
Function | Range / Description |
DC voltage | 0 to 1000V |
AC voltage,TRMS | 2.5 mV to 760V |
DC current | 0 to 10A |
AC current,TRMS | 20 μA to 10A |
Resistance | 0 to 50MΩ |
Capacitance | 0.5nF to 110.00mF |
Diode | About 3.5V |
TC Test | K |
RTD Test | Pt100 |
Frequency | TYPE A:3Hz to 100kHz,TYPE B:3Hz to 10kHz |
Data storage | 50000 groups |
Function | Range | Measurement Range | Resolution | Accuracy |
DCmV | 50mV | -55.000mV~55.000mV | 0.001mV | 0.1%+10 |
500mV | -550.00mV~550.00mV | 0.01mV | 0.025%+5 | |
DCV | 5V | -5.5000V~5.5000V | 0.0001V | 0.025%+5 |
50V | -55.000V~55.000V | 0.001V | 0.03%+5 | |
500V | -550.00V~550.00V | 0.01V | 0.1%+5 | |
1000V | -1000.0V~1000.0V | 0.1V | 0.1%+5 |
Function | Range | Measurement Range | Resolution | Accuracy |
DCI | 500uA | -550.00uA~550.00uA | 0.01uA | 0.15%+20 |
5000uA | -5500.0uA~5500.0uA | 0.1uA | 0.15%+10 | |
50mA | -55.000mA~55.000mA | 0.001mA | 0.15%+20 | |
500mA | -500.00mA~500.00mA | 0.01mA | 0.15%+10 | |
5A | -5.5000A~5.5000A | 0.0001A | 0.5%+10 | |
10A | -10.000A~10.000A | 0.001A | 0.5%+10 |
Measurement Function | Range | Measurement Range | Resolution | Accuracy | ||||||
20Hz~45Hz | 45Hz~65Hz | 65Hz~400Hz | 400Hz~1KHz | 1KHz~10KHz | 10KHz~20KHz | 20KHz~50KHz | ||||
ACmV | 50mV | 0~55.000mV | 0.001mV | 2%+80 | 0.7%+30 | 0.7%+30 | 2%+30 | 3%+60 | 5%+100 | |
500mV | 0~550.00mV | 0.01mV | 2%+80 | 0.7%+30 | 0.7%+30 | 2%+30 | 3%+60 | 5%+100 | ||
ACV | 5V | 0~5.5000V | 0.0001V | 2%+80 | 0.5%+30 | 0.7%+30 | 2%+30 | 3%+60 | 5%+100 | |
50V | 0~55.000V | 0.001V | 2%+80 | 0.5%+30 | 0.7%+30 | 2%+30 | 3%+60 | 5%+100 | ||
500V | 0~550.00V | 0.01V | 2%+80 | 0.5%+30 | 0.7%+30 | 2%+30 | ----- | ----- | ----- | |
1000V | 0~760.0V | 0.1V | 2%+80 | 0.5%+30 | 0.7%+30 | 2%+30 | ----- | ----- | ----- | |
ACA | 10A | 0~10.000A | 0.001A | 2%+40 | 1.5%+20 | 3%+80 | ----- | ----- | ||
5A | 0~5.5000A | 0.0001A | 2%+40 | 1%+20 | 3%+80 | ----- | ----- | |||
ACmA | 50mA | 0~55.000mA | 0.001mA | 1.5%+80 | 0.5%+30 | 1%+30 | 2%+603 | 2%+60 | ||
500mA | 0~500.00mA | 0.01 mA | 1.5%+80 | 0.5%+30 | 1%+30 | 2%+603 | 2%+60 | |||
ACuA | 500uA | 0~550.00uA | 0.01uA | 1.5%+80 | 0.5%+30 | 1%+30 | 2%+603 | 2%+60 | ||
5000uA | 0~5500.0uA | 0.1uA | 1.5%+80 | 0.5%+30 | 1%+30 | 2%+603 | 2%+60 | |||
1.<20kHz is adapt to 5~110% range, ≥20kHz is adapt to 10~110% range. 2.1000V and 10A range is adapt to 10~110% range. 3.50mV、500mV range is 30Hz~50KHz. 4.ACmA、ACuA range is 20Hz~30KHz. |
Measurement Function | Range | Measurement Range | Resolution | Accuracy | ||
20Hz~45Hz | 45Hz~400Hz | 400Hz~1KHz | ||||
ACV | 5V | 0~5.5000V | 0.0001V | 2%+80 | 0.5%+30 | 3%+30 |
50V | 0~55.000V | 0.001V | 2%+80 | 0.5%+30 | 2%+30 | |
500V | 0~550.00V | 0.01V | 2%+80 | 0.5%+30 | 2%+30 | |
1000V | 0~760.0V | 0.1V | 2%+80 | 0.5%+30 | 2%+30 | |
ACmV | 50mV | 0~55.000mV | 0.001mV | 2%+80 | 0.5%+30 | 0.7%+30 |
500mV | 0~550.00mV | 0.01mV | 2%+80 | 0.5%+30 | 0.7%+30 | |
ACA | 10A | 0~10.000A | 0.001A | 2%+40 | 1.5%+20 | |
5A | 0~5.5000A | 0.0001A | 2%+40 | 1%+20 | ||
ACmA | 50mA | 0~55.000mA | 0.001mA | 1.5%+80 | 0.7%+30 | |
500mA | 0~500.00mA | 0.01mA | 1.5%+80 | 0.7%+30 | ||
ACuA | 500uA | 0~550.00uA | 0.01uA | 1.5%+80 | 0.7%+30 | |
5000uA | 0~5500.0uA | 0.1uA | 1.5%+80 | 0.7%+30 |
Function | Range | Measurement Range | Resolution | Accuracy |
OHM | 500Ω | 0~550.00Ω | 0.01Ω | 0.05%+10 |
5KΩ | 0~5.5000KΩ | 0.0001KΩ | 0.05%+5 | |
50KΩ | 0~55.000KΩ | 0.001KΩ | 0.05%+5 | |
500KΩ | 0~550.00KΩ | 0.01KΩ | 0.05%+5 | |
5MΩ | 0~5.5000MΩ | 0.0001MΩ | 0.2%+5 | |
50MΩ | 0~55.000MΩ | 0.001MΩ | 1%+10 | |
1. Use the Null mode to zero the error readings. 2. The accuracy does not include the error caused by the lead resistance. |
Function | Range | Measurement Range | Resolution | Accuracy |
CAP | 10nF | 0~11.00nF | 0.01nF | 5%+50 |
100nF | 0~110.0nF | 0.1nF | 5%+5 | |
1000nF | 0~1100nF | 1nF | 5%+5 | |
10µF | 0~11.00µF | 0.01uF | 5%+5 | |
100µF | 0~110.0µF | 0.1uF | 5%+5 | |
1000µF | 0~1100µF | 1uF | 5%+50 | |
10mF | 0~11.00mF | 0.01mF | 5%+50 | |
100mF | 0~110.0mF | 0.1mF | 5%+50 | |
1. Use the Null mode to zero the residual readings. 2. For readings high than 10mF, the last digit of the reading is invalid. |
Function | Range | Measurement Range | Resolution | Accuracy | Sensitivity |
Frequency | 10Hz | 0~10.000Hz | 0.001Hz | 0.02%+4 | 500mV_RMS |
100Hz | 0~100.00Hz | 0.01Hz | 0.02%+4 | ||
1000Hz | 0~1000.0Hz | 0.1Hz | 0.02%+4 | ||
10kHz | 0~10.000kHz | 0.001kHz | 0.02%+4 | ||
100kHz | 0~100.00kHz | 0.01kHz | 0.02%+4 | ||
DUTY | 10%~90% | 0.1% | 0.3%/KHz+0.3% | ||
1. For signal reading lower than 3Hz, the reading is Zero. 2. TYPE A:3Hz to 100kHz,TYPE B:3Hz to 10kHz |
Function | Range | Measurement Range | Resolution | Sensitivity |
DIODE | 2V | 0.0001V | 1%+10 |
Function | Range | Measurement Range | Resolution | Sensitivity |
Continuity Test | 600Ω | 0.1Ω | ≤30ΩBB |
Graduation No. | Measurement Range | Resolution | Accuracy |
K | -200.0°C~1372.0°C | 0.1°C | 0.5%+0.5°C |
-328.0°F~2501.6°F | 0.1°F | 0.5%+1°F | |
1. By using ITS-90 temperature scale . 2. The accuracy does not include the error of internal temperature compensation. |
Graduation No. | Measurement Range | Resolution | Accuracy |
Pt100 | -200.0°C~800.0°C | 0.1°C | |
-328.0°F~1472.0°F | 0.1°F | ||
1. By using ITS-90 temperature scale 2. Attached lead resistance is excluded |
Measurement Function | Range | Overload Voltage |
Current | 10.000A | 0.04V/A |
5.0000A | 0.04V/A | |
500.00mA | 1.8mV/mA | |
50.000mA | 1.8mV/mA | |
5000.0μA | 103uV/μA | |
500.00μA | 103uV/μA |
Function Position | Input Impedance (Standard Value) | |||||
V | 10MΩ,<100pF | |||||
mV | >2.5GΩ | |||||
Common Mode Rejection Ratio | Normal Mode Rejection | |||||
DCV、DCmV | 80dB(dc to 50Hz / 60Hz/1KΩ) | 40dB(50Hz / 60Hz) | ||||
ACV、ACmV | 80 dB(dc to 50Hz / 60Hz/1KΩ) | |||||
Open Circuit Test Voltage | Full-scale Voltage | |||||
ohm | 2.0V | 2.0V | ||||
Diode | < 3.5V | 2.0V | ||||
Continuity | < 1.5V | 500mV | ||||
Typical short-circuit current | ||||||
ohm | 500Ω | 5KΩ | 50KΩ | 500KΩ | 5MΩ | 50MΩ |
0.8mA | 0.2mA | 20μA | 2μA | 0.2μA | <0.1μA | |
Diode | 0.2mA(Typical Value) |